Max Mobility / Permobil Announces Worldwide Expanded Recall and Removal of all SpeedControl Dials from the Market Due to Intermittent Cable Connection Concerns of the Dial with the SmartDrive MX2+ Motor. Users Must Use an Alternate Control Method for their SmartDrive MX2+ Power Assist Device.
- Date
- Aug 12, 2025
- Company
- Max Mobility, LLC.
- Brand
- Max Mobility/Permobil
- Product
- Speed Control Dial Component
- Reason
- Due to potential safety and performance concerns
- Hazard
- Other
- Source
- FDA
How to identify the product
- This voluntary action applies to all SpeedControl Dials manufactured and distributed between the dates of April 25, 2022 to July 08, 2025.
- A Class I recall (Z-1116-2025, Z-1117-2025), RES # 96086 was originally initiated on December 20, 2024 to address a material change with the printed circuit board assembly (PCBA) in SpeedControl Dials manufactured and distributed between the dates of August 17, 2023 and November 21, 2024.
- The recall was later expanded on March 28, 2025, and included all units manufactured and distributed from August 17, 2023 up to March 10, 2025.
- Max Mobility / Permobil Issues Nationwide Recall of SmartDrive Speed Control Dial Due to the Motor Being Unresponsive to the User | FDA
- Max Mobility / Permobil Expands Nationwide Recall of SmartDrive Speed Control Dial Due to the Motor Being Unresponsive to the User | FDA
- This design issue has been part of the SpeedControl Dial since it was first introduced to the market in April 2022 and impacts all SpeedControl Dials manufactured and distributed, which occurred between April 25, 2022 to July 08, 2025, if there is an intermittent electrical connection between the SpeedControl Dial and SmartDrive MX2+ motor, the following performance issues could occur: