Medline Industries, LP Issues Nationwide Recall of Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits
- Date
- Aug 5, 2026
- Company
- Medline Industries, LP
- Brand
- Medline, Teleflex
- Product
- Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits
- Reason
- Due to heated wire connector (pigtail) at the Neptune humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage
- Hazard
- Fire / burn
- Source
- FDA
How to identify the product
- On July 13, 2026, Medline Industries, LP initiated a nationwide recall of Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits.
- To date, reported events have occurred infrequently relative to the number of devices distributed, with the observed occurrence rate of 0.011% across the distributed population.
- Consumers should not return product to Medline.
- Recalled Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits were manufactured on or before July 8, 2026 and distributed from June 28, 2021, to July 31, 2026.